Medtronic Announces Worldwide Voluntary Field Corrective Action for Newport™ HT70 and Newport™ HT70 Plus Ventilators

Medtronic (NYSE: MDT) is notifying customers worldwide of a voluntary field corrective action for all its Newport HT70 and Newport HT70 Plus ventilators manufactured since 2010. The voluntary field corrective action is being conducted following reports that the ventilator may reset spontaneously during normal operation, without an accompanying alarm. The reported incidence of this condition is approximately one (1) reset in every seven million hours of ventilation. Following the reset, the ventilator enters standby mode and will not resume ventilation without intervention. In the event of the rare occurrence of a reset, healthcare professionals and/or caregivers are required to transfer the patient to another ventilator.